Benchtop conductivity meters for laboratory analysis.

FEATURES:
- Reliable and precise measurements
- Fast and easy GLP documentation
- Auto sensor recognition for Lab 970
Benchtop Conductivity Meters Product Description
The Lab 960 benchtop conductivity meter was designed to perfectly match SI Analytics world class electrodes. The benchtop conductivity meter offers the highest level of GLP documentation. The benchtop conductivity meters offer GLP compliant protocols and up to 800 data sets with intervals from five seconds to 60 minutes storing both time and identification information. The benchtop conductivity meters also feature data transmission via a conventional RS232 port or USB interface.
Lab 970 benchtop conductivity meter
The Lab 970 benchtop conductivity meters include all the features of the Lab 960 conductivity meter except for data storage. The Lab 970 benchtop conductivity meters send the data directly to your computer via their USB or RS232 ports. This allows the benchtop conductivity meters to be easily handled and integrated into automated systems such as LIMS. The Lab 970 benchtop conductivity meters are ideally suited for measurements according to USP 28. This means that the benchtop conductivity meters automatically recognize which conductivity electrode is attached. The conductivity electrodes use the latest microtechnology to enable the storage of calibration data in the electrode to be read by the benchtop conductivity meter. For each measurement the conductivity electrode logs into the benchtop conductivity meter with its ID (type and serial number) and sends its calibration data to assure that only the sensor’s specific data is used to calculate the measured value. Complete IQ and OQ documentation is available for the benchtop conductivity meters.
Handheld Conductivity Meters Specifications
Conductivity:
Range: 0.000 µS/cm to 500 mS/cm
TDS Range: 0.4 to 1.0
Salinity: Measured according to Natural Sea Water Scale (UNESCO 1966b)
Accuracy (±1 digit): 0.5 % of measured value
Calibrated Cell Constant: 0.450 to 0.500 cm -1; 0.585 to 0.715 cm -1; 0.800 to 1.200 cm -1
Adjustable Cell Constant: 0.250 to 2.500 cm -1 and 0.090 to 0.110 cm -1
Fixed Cell Constant: 0.010 cm -1
Temperature Compensation: nLF/Lin (0.001 to 3.000 %/K)/selectable
Temperature Compensation Purity Water: Yes
Reference Temperature: 68 °C or 77 °F (20 °C or 25 °C) Selectable
Temperature:
Range: 23 to 248 °F (-5.0 to 120.0 °C)
Accuracy (±1 digit): ±0.2 °F (±0.1 °C)
Selectable °F/°C units: Yes
Automatic switch to manual temperature input when no temperature sensor is connected: Yes
Mechanical and electrical
Display: LCD, 3×2.4 inch (75×60 mm)
Data Storage: Lab 960 only – 800 data sets with intervals from 5 seconds to 60 minutes
USB (slave) and RS232 interface: Yes
Lower Casing: Plastic
Keypad: Plastic foil with tactile response
Power Source: External 100-240VAC, 50/60 Hz or 9V battery
Automatic power shutoff (battery): Yes, adjustable 10 minutes to 24 hrs
Dimensions (WxHxD): 9.45×7.48×3.15in (240x190x80mm)
Weight: 2.2 lb (1.0 kgs)
Compliance: CE, cETLus
Safety: protection class III
Climate class: 2 (VDI/VDE 3540)
Warranty: 3 years
LAB 960 BENCHTOP CONDUCTIVITY METERS AND KITS
Lab 960 Benchtop Conductivity Meter Set(PN# 285204180)
Measuring ranges 0.000 µS/cm to 500 mS/cm, salinity, total dissolved solids (TDS), temp., RS232 C and USB (slave) interface, micropr., storage for 800 data, GLP conform. Including cover Z 880, stand, power supply Z 850, and conductivity cell LF 413 T.
Lab 960 Benchtop Conductivity Meter (PN# 285204170)
Meter only, measuring ranges 0.000 µS/cm to 500 mS/cm, salinity, total dissolved solids (TDS), temp., RS232 C and USB (slave) interface, microprocessor, data storage for 800 data, GLP conform.
LAB 970 BENCHTOP CONDUCTIVITY METER KITS
Lab 970 Benchtop Conductivity Meter Set(PN# 285204190)
Sensor recognition. Measuring ranges 0.000 µS/cm…500 mS/cm, salinity, total dissolved solids (TDS), temp., RS232 C and USB (slave) interface, micropr., GLP conform. Including cover Z 880, stand, power supply Z 850, and conductivity cell LF 413 T ID.
BENCHTOP CONDUCTIVITY METER ELECTRODES
LF 413T Conductivity Electrode (PN# 285106172)
4 pole cell, plastic shaft, 4.9 ft (1.5 m) cable with 8 pole plug, sensor material graphite, cell constant 0.475 cm-1, temp.-sensor NTC 30 kOhm, length 4.7 in (120 mm), 0.6 inch Ø (15.3 mm Ø), 23 to 176 °F (-5 to +80 °C)
LF 413T-ID Conductivity Electrode (PN# 285130310)
4 pole cell, plastic shaft, 4.9 ft (1.5 m) cable with 8 pole plug and recognition, sensor material graphite, cell constant 0.475 cm-1, temp.-sensor NTC 30 kOhm, length 4.7 in (120 mm), 0.6 inch Ø (15.3 mm Ø), 23 to 176 °F (-5 to +80 °C)
Conductivity Electrodes
Conductivity electrodes for various applications.
BENCHTOP CONDUCTIVITY METER CALIBRATION SOLUTION
LF 990 147 µS/cm Conductivity Solution, 3×6 ampoules (PN# 285126503)
Potassium chloride 0.001 mol/l
LF 991 1.41 mS/cm Conductivity Solution, 3×6 ampoules (PN# 285126511)
3×6 20 mL ampoules, test solution potassium chloride 0.01 mol/l (1.41 mS/cm)
LF 992 12.9 mS/cm Conductivity Solution, 3×6 ampoules (PN# 285126528)
3×6 20 mL ampoules, test solution potassium chloride 0.1 mol/l (12.9 mS/cm)
LF 995 1.41/12.9/112 mS/cm Conductivity Solution, 3×6 ampoules (PN# 285126293)
Potassium chloride 0.01/0.1/1 mol/l
BENCHTOP CONDUCTIVITY METER ACCESSORIES
Z 880 Cover for Lab 960 and 970 benchtop conductivity meters (PN# 285201550)
Z 865 Stand set S4D (PN# 285201520)
Includes arm and electrode holder for the Lab 960 and 970 benchtop conductivity meters.
Z 850 Power Supply (PN# 285204889)
Universal power supply unit, 120/230V, for use with Lab 960 and 970 benchtop conductivity meters.
RELATED PRODUCTS
EC30A Conductivity Tester
Includes meter and conductivity cell, batteries, and instruction manual.
EC300ACC Field Conductivity Meter Kit
Includes field conductivity meter, cable with attached conductivity/temperature probe, and hard sided case.
EC500 Pocket pH/Conductivity Tester
Includes electrode, protective sensor cap, sample cup with cap, four 1.5V SR44W batteries, and a 48 inch (1.2m) neckstrap.
EC510 Pocket pH/Conductivity Tester Kit
Includes an EC500 portable conductivity meter with sensor, 84uS, 1413uS, 12880uS calibration standards, pH buffer pouches (1 each of 4, 7, 10pH plus rinse solution), weighted base, protective sensor cap, 3 sample cups, four 1.5V SR44W batteries, carrying case, and a 48 inch (1.2m) neckstrap.
2CA102 Cond 3110 Set Handheld Conductivity Meter Set
Includes meter, TetraCon 325-3 probe, case & accessories.
WQ-Cond Conductivity Sensor
4-20mA output. Includes 25 ft of cable.